Sample Preparation: The sample was dried in solid state to protect the equipment. Approximately 2 mg of the sample was weighed directly into a Tzero hermetic aluminum pan to avoid material loss. A lid (indention facing down) was placed on the pan and seated flush in a blue holder. The pan was sealed using a press. A reference pan (empty and sealed) was prepared in the same manner. DSC Procedure: Both sample and reference pans were placed in their respective tray slots, with slot numbers noted. Cooler power was turned on in event mode. TA Universal Analysis software was launched: Under the Control tab, the cooler was started and set to standby temperature. Nitrogen gas was turned on and regulated at 20 psi with a sample purge flow of 50 mL/min. In the program panel, sample information was entered: name, weight, slot numbers, and data storage location. Under the Procedure tab, experimental parameters (temperatures, ramp rates, etc.) were defined and saved. In the Experiment panel, runs were arranged and started in the desired order. After completion, the cooling system shut down automatically; the nitrogen tank was turned off manually.